| Intended Use: | For Self Monitoring of Blood Glucose |
| Method: | Amperometric Glucose Biosensor |
| Physical Characteristic: | 1.38” – 1.5” x .236” x .031” |
| Material: Base Material: | PET |
| Electrode: | Carbon |
| Chemical Components: | Enzyme: Glucose Oxidose |
| Mediator: Potassium Ferrycyanide | |
| Others: Non-Reactive ingredient | |
| Function: | Sample Type: Capillary Whole Blood; Finger Tip or AST |
| Sample Volume: Tiny Blood Sample | |
| Sampling Method: Capillary | |
| Reaction Time: Rapid Test Time | |
| Performance: | Linear Range: 20 ~ 600 mg/dL (1.1 ~ 33.3 mml/L) |
| Precision: < 45% with respect to control solution | |
| Accuracy: > 95% fall in zone A & B (Blood EGA Test) | |
| Stability: | Operation: 10°C ~ 40°C, RH < 85% |
| Storage: 4°C ~ 40°C, RH < 85% | |
| Validity: | 90 days from vial being opened, or 18 months from production date |
WARNING!
- Do Not Use Expired Strip
- These Strips are Only Used with Advocate Redi-Code Glucose Meters
- Close Vial Immediately Every Time a Strip is Taken Out
- Cannot Put Fluoride into Blood as the Preservative
- Not recommend to do this Test on New Born Babies
- High Density of Vitamin C or Uricacid may Influence Glucose Value
- Hematocrit above 60% may Cause Lower Glucose Value
- Hematocrit Lower than 20% may cause Higher Glucose Value